High-Temperature Furnaces with SiC Rod Heating up to 1600 °C

Heating times of 25 - 30 minutes can be achieved, depending on the furnace model and the conditions of use.

Process Control

Function and equipment

LHTCT01_16_3.png
LHTC08_16
Begasungssystem
Ofenraum_SiC

These powerful laboratory muffle furnaces are available for temperatures up to 1550 °C or 1600 °C. The durability of the SiC rods in periodic use, in combination with their high heating speed, make these high-temperature furnaces to all-rounders in the laboratory. Heating times of 25 - 30 minutes can be achieved, depending on the furnace model and the conditions of use.

Standard Equipment

  • Tmax 1550 °C or 1600 °C
  • Working temperature 1500 °C, increased wear and tear must be expected in case of working at higher temperatures
  • Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
  • Exclusive use of insulation materials without categorization according to EC Regulation No. 1272/2008 (CLP)
  • Optional flap door (LHTC) which can be used as work platform or lift door (LHTCT) with hot surface facing away from the operator (High-temperature furnace LHTCT 01/16 only with lift door)
  • Switching system with solid-state-relays, power tuned to the SiC rods
  • Easy replacement of heating rods
  • Adjustable air inlet opening, exhaust air opening in the roof
  • Controller C550 (10 programs with each 20 segments)
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive

 

Additional equipment

Additional Equipment

  • Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
  • Protective gas connection to purge with non-flammable protective or reaction gases, not gas tight
  • Manual or automatic gas supply system
  • Process control and documentation via VCD software package for monitoring, documentation and control

Technical specifications

Model Tmax Inner dimensions in mm Volume Outer dimensions1 in mm Connected Electrical Weight Heating time
  in °C w d h in l W D H2 load in kW connection* in kg in min3
LHTCT 01/16 1550 110 120 120 1.5 340 300 460 3.5 1-phase 18 30
LHTC(T) 03/16 1600 120 210 120 3.0 415 545 490 10.0 3-phase4 30 30
LHTC(T) 08/16 1600 170 290 170 8.0 490 625 540 15.5 3-phase 40 25
1External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.*Please see page 75 for more information about supply voltage
2Plus maximum 240 mm for models LHTCT when open
3Heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE rsp. 400 V 3/N/PE)
4Heating only between two phases