Ashing Furnaces up to 1100 °C

A special fresh-air and exhaust air system ensures that the air is replaced 6 times per minute so that there is always sufficient oxygen for the ashing process.

Process Control
Ashing furnace LVT 3/11

Function and equipment

LV5_11.png
LVT3_11.png
LVT9_11.png
LV_Luftfuehrung.png
LV_Ofeninnenraum.png
L_Thermoelementdurchfuehrung_fmt.jpg
LV_Beschickungsgestell_fmt.jpg

Ashing furnace LV(T) .. 11 is designed especially for ashing processes to 1050 °C in the laboratory. Applications include determining loss on ignition, ashing food and plastics for subsequent substance analysis. A special fresh-air and exhaust air system ensures that the air is replaced 6 times per minute so that there is always sufficient oxygen for the ashing process. Incoming air passes the furnace heating and is pre-heated to ensure good temperature uniformity.

Standard Equipment

  • Tmax 1100 °C
  • Heating from two sides
  • Ceramic heating plates with integral heating element which is safeguarded, and easy to replace
  • Air exchange of more than 6 times per minute
  • Good temperature uniformity due to preheating of incoming air, temperature uniformity according to DIN 17052-1 to +/− 10 °C in the defined empty work area (from 550  °C)
  • Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
  • Exclusive use of insulation materials without categorization according to EC Regulation No. 1272/2008 (CLP)
  • Suitable for many standardized ashing processes according to ISO, ASTM, EN, and DIN
  • Optional flap door (LV) which can be used as work platform or lift door (LVT) with hot surface facing away from the operator
  • Controller with touch operation B510 (5 programs with each 4 segments)
  • Solid state relays provide for lownoise operation
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive

Additional equipment

Additional Equipment

  • Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
  • Port for thermocouple in the rear wall or in the furnace door
  • Charging rack with closed or perforated trays for loading the furnace in two levels incl. holder for inserting/removing the trays up to a max. temperature of 800 °C and a max. loading weight per layer of 2 kg for the LV(T) 9/11 respectively 3 kg for the LV(T) 15/11
  • Process control and documentation via VCD software package for monitoring, documentation and control

Technical specifications

Model

Tmax

Inner dimensions in mm

Volume

Outer dimensions2 in mm

Max. weight of hydrocarbons

Max. evaporation rate

Max. connected
load

Electrical

Weight

Heating time

Flap door

in °C1

w

d

h

in l

W

D

H3

in g

g/min

in kW

connection

in kg

in min4

LV 3/11

1100

180

150

120

3

345

390

810

5

0.1

1.3

1-phase

20

45

LV 5/11

1100

205

170

130

5

385

415

810

10

0.2

2.6

1-phase

29

55

LV 9/11

1100

235

240

170

9

415

485

865

15

0.3

3.3

1-phase

36

70

LV 15/11

1100

230

340

170

15

415

590

865

25

0.3

3.6

1-phase

44

80


 

Model

Tmax

Inner dimensions in mm

Volume

Outer dimensions2 in mm

Max. weight of hydrocarbons

Max. evaporation rate

Max. connected load

Electrical

Weight

Heating time

Lift door

in °C1

w

d

h

in l

W

D

H3

in g

g/min

in kW

connection

in kg

in min4

LVT 3/11

1100

180

150

120

3

345

390

810

5

0.1

1.3

1-phase

20

45

LVT 5/11

1100

205

170

130

5

385

415

810

10

0.2

2.6

1-phase

29

55

LVT 9/11

1100

235

240

170

9

415

485

865

15

0.3

3.3

1-phase

36

70

LVT 15/11

1100

230

340

170

15

415

590

865

25

0.3

3.6

1-phase

44

80

1Recommended working temperature for processes with longer dwell times is 1000 °C
2External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.
3Including exhaust tube (Ø 80 mm)
4Approx. heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE)


 

Charging rack for model

Articel No.

Tmax

Outer dimensions in mm

Number of levels

Dimensions shelf (level 1) in mm

Max. weight per level

  

in °C

W

D

H

 

W

D

H

in kg

LV(T) 9/11

6000079693

800

215

218,5

95

2

202

202

47

2

LV(T) 15/11

6000078459

800

215

318,5

95

2

202

302

47

3


 

Request for quotation

Anfrage

CAPTCHA