Bысокотемпературные печи с нагревательными элементами из MoSi₂ до 1800 °C
Благодаря превосходно обработанным высококачественным использованным материалам в сочетании с удобством управления эти печи являются незаменимыми помощниками в исследованиях и лабораторных опытах.
Функции и оборудование
High-temperature furnace LHT 02/17
Традиционная агломерационная печь LHT 01/17 D
High-temperature furnace LHT 03/17 D
High-temperature furnace LHT 08/18
Резервуар для садок с крышкой
Furnace chamber with high-quality fiber materials and heating elements made of molybdenum disilicide on both sides
Пример регулируемого ограничителя температуры
Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages. The first-class workmanship using high-quality materials, combined with ease of operation, make these furnaces all-rounders in research and the laboratory. These high-temperature furnaces are also perfectly suited for the sintering of technical ceramics, such as zirconium oxide dental bridges.
Standard Equipment
- Tmax 1600 °C, 1750 °C, or 1800 °C
- Recommended working temperature 1750 °C (for models LHT ../18), increased wear and tear must be expected in case of working at higher temperatures
- High-quality heating elements made of molybdenum disilicide offer very good protection against chemical interaction between charge and heating elements
- Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
- Exclusive use of insulation materials without categorization according to EC Regulation No. 1272/2008 (CLP)
- Adjustable air inlet opening
- Exhaust air opening in the roof
- Thermocouple type B or type S (LHT ../17 D)
- Controller with touch operation P580 (50 programs with each 40 segments)
- Defined application within the constraints of the operating instructions
- NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive
Дополнительное оборудование
Additional Equipment
- Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
- Port for thermocouple in the furnace roof
- Protective gas connection to purge with non-flammable process, not gas tight
Manual or automatic gas supply system - Stackable saggars for loading in up to two or three levels, depending on model,
- Process control and documentation via VCD software package for monitoring, documentation and control
Спецификации
| Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions1 in mm | Connected | Electrical | Weight | Heating time | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| in °C | w | d | h | in l | W | D | H2 | load5 in kW | connection | in kg | in min3 | |
| LHT 02/16 | 1600 | 130 | 145 | 130 | 2 | 430 | 450 | 570+325 | 3.4 | 1-phase | 33 | 28 |
| LHT 04/16 | 1600 | 160 | 175 | 160 | 4 | 450 | 475 | 610+335 | 3.4 | 1-phase | 39 | 50 |
| LHT 08/16 | 1600 | 200 | 200 | 200 | 8 | 500 | 500 | 650+370 | 6.7 | 3-phase4 | 47 | 33 |
| LHT 01/17 D | 1650 | 110 | 120 | 120 | 1 | 385 | 425 | 525+195 | 2.7 | 1-phase | 28 | 27 |
| LHT 03/17 D | 1650 | 135 | 135 | 200 | 4 | 412 | 450 | 595+300 | 2.7 | 1-phase | 38 | 57 |
| LHT 02/17 | 1750 | 130 | 145 | 130 | 2 | 430 | 450 | 570+325 | 3.4 | 1-phase | 33 | 46 |
| LHT 04/17 | 1750 | 160 | 175 | 160 | 4 | 450 | 475 | 610+335 | 3.4 | 1-phase | 39 | 90 |
| LHT 08/17 | 1750 | 200 | 200 | 200 | 8 | 500 | 500 | 650+370 | 6.7 | 3-phase4 | 47 | 50 |
| LHT 02/18 | 1800 | 130 | 145 | 130 | 2 | 430 | 450 | 570+325 | 3.4 | 1-phase | 33 | 56 |
| LHT 04/18 | 1800 | 160 | 175 | 160 | 4 | 450 | 475 | 610+335 | 3.4 | 1-phase | 39 | 106 |
| LHT 08/18 | 1800 | 200 | 200 | 200 | 8 | 500 | 500 | 650+370 | 6.7 | 3-phase4 | 47 | 60 |
| 1External dimensions vary when furnace is equipped with additional equipment. Dimensions on request. 2Including opened lift door4Heating only between two phases 3Heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE rsp. 400 V 3/N/PE) 5The connected load refers to the standard furnace and may increase for a furnace with additional equipment. For furnaces with connection options for multi-range voltages, the connected load applies to the highest permissible connected voltage. | ||||||||||||
Запрос коммерческого предложения
Другие продукты
каталог
Полный обзор нашего обширного ассортимента печей, сушильных шкафов и принадлежностей для лабораторных помещений вы найдете в нашем каталоге «Лаборатория».