Высокотемпературные печи с весами для определения потери при прокаливании и термогравиметрического анализа (ТГА) до 1750 °C

В комплект поставки входит опорный каркас под печь, керамический плунжер с плитой во внутреннем пространстве печи, прецизионные весы и пакет программного обеспечения.

Управление процессами
High-temperature furnace LHT 04/17 SW with scale for measuring weight reduction during annealing

Функции и оборудование

LHT04-17SW.png
LHT-SW-Waage.png
LHT_Mosi2_Heizelemente
Grafische_Darstellung.png

These high-temperature furnaces were specially developed to determine combustion loss during annealing and for thermogravimetric analysis (TGA) in the lab. The complete system consists of the high-temperature furnace for 1600 °C or 1750 °C, a table frame, precision scale with feedthroughs into the furnace and powerful software for recording both the temperature curve and the weight loss over time.

Standard Equipment

  • Defined application within the constraints of the operating instructions
  • Tmax 1600 °C or 1750 °C
  • High-quality molybdenum disilicide heating elements
  • Dual shell ventilated housing made of textured stainless steel sheets for low surface temperature and high stability
  • Exclusive use of insulation materials without categorization according to EC Regulation No 1272/2008 (CLP). This explicitly means that alumino silicate wool, also known as “refractory ceramic fiber” (RCF), which is classified and possibly carcinogenic, is not used.
  • Adjustable air inlet
  • Exhaust air opening in the roof
  • Type B thermocouple
  • Delivery includes base, ceramic plunger with base plate in the furnace lining, precision scale and software package
  • 4 scales available for different maximum weights and scaling ranges
  • Process control and documentation for temperature and combustion loss via VCD software package for monitoring, documentation and control

Спецификации

ModelTmaxInner dimensions in mmVolumeOuter dimensions1 in mmConnectedElectricalWeightHeating time
 in °Cwdhin lWDHload4 in kWconnectionin kgin min2
LHT 04/16 SW1600150150150465537089053-phase38525
LHT 04/17 SW1750150150150465537089053-phase38530
1External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.
2Heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE rsp. 400 V 3/N/PE)
3Heating only between two phases
4The connected load refers to the standard furnace and may increase for a furnace with additional equipment. For furnaces with connection options for multi-range voltages, the connected load applies to the highest permissible connected voltage.