Installation and Exhaust Air System

When exhaust gas pipings are installed it is always necessary that a local ventilation technician lays out the system in accordance to the real environment. 
 

Process Control

Function and equipment

Mindestabstand.png
Ableitungsmoeglichkeiten.png

Installation
When the kiln is being installed, it is important that there is a safety gap of 0.5 m between the kiln and flammable materials on all sides and 1.0 m to the ceiling. If the ceiling is lower, heat-resistant insulation must be used for protection. The kiln must be placed on a non-flammable surface (fire safety class A DIN 4102 – Example: concrete, tiles, glass, aluminum or steel). The floor must be level so that the kiln can stand upright. Kiln and switchgear are not designed to be used outdoors.

Exhaust air system
When exhaust gases are generated during the process it is mandatory to guide them outside in an adequate way. The relevant operating instructions must be always taken into consideration. When exhaust gas pipings are installed it is always necessary that a local ventilation technician lays out the system in accordance to the real environment.

There are different possibilities to guide the exhaust gases out. In many cases the furnace is positioned under a laboratory extraction provided by the customer. In these cases the use of an exhaust vent is recommended just to guide the gases upwards.

For this purpose metal exhaust gas pipes with NW 80 to NW 120 can be used. They must be installed continuously rising and fastened to the wall or ceiling. Center the pipe over the furnace vent (for models with vent fan or catalytic converter, NW 120 is necessary. The exhaust gas pipe must not be installed with a tight fit to the furnace vent pipe since this would prevent any bypass effect. This is necessary so that not too much fresh air is sucked in by the furnace. 