Function and equipment
Gas Supply System 4 for Hydrogen Applications in Tube Furnaces from Room Temperature
Gas supply system 4 allows operation with a hydrogen atmosphere starting at ambient temperature. During hydrogen operation, a pressure of approx. 30 mbar is ensured in the working tube. At the gas outlet the hydrogen is burnt off in an exhaust gas torch. Equipped with a safety PLC control system, pre-purging, hydrogen inlet, operation, fault monitoring and purging at the end of the process are carried out automatically (with at least five times the volume of the tube). If a malfunction occurs, the tube is immediately purged with nitrogen and the system is automatically switched to a safe status.
Standard Equipment
- Available for R, RHTC, RHTH, RHTV, RSH, RSV, RSRB and RSRC tube furnaces
- Gas panel for hydrogen and nitrogen
- Automatic segment-related switching on/off by a magnetic valve
- Control via safety PLC control system with touch panel
- Exhaust gas torch with temperature monitoring
- Over-temperature limiter with digital display as over-temperature protection for the furnace and charge
- Excess pressure monitoring
- Emergency purge container for nitrogen
Additional equipment
Additional Equipment
- Additional gas panels for further non-flammable gases
- Operation with other flammable gases
- Gassing via program-related controllable mass flow controllers
- Bottle pressure reducer for use with bottled gas
- Air-water heat exchanger for closed loop water circuit (apart from RSRB and RSRC)
Technical specifications
Model | Gas supply system | |||||
---|---|---|---|---|---|---|
1 | 15 | 2 | 25 | 26 | 4 | |
RD | ● | |||||
R | ● | ● | ● | ● | ||
RT | ● | |||||
RHTC | ● | ● | ● | |||
RHTH | ● | ● | ||||
RHTV | ● | ● | ||||
RSH | ● | ● | ● | ● | ||
RSV | ● | ● | ● | ● | ||
RSRB | ● | ● | ||||
RSRC | ● | ● |
Request for quotation
Further products
Catalog
A complete overview of our extensive range of furnaces, ovens and accessories for the laboratory can be found in our “Laboratory” catalog.